DeepArrhythmia: Segment-Contextualized ECG Arrhythmia Classification via Selective Evidence Acquisition

Update NeurIPS 2026 submission i8JkJYOjY6: 9 notes

@openreview[bot] · 2026-11-24 15:54 UTC · 0bfbdac659b7 ·

Evidence

reviews/openreview/i8jkjyojy6/notes/5vfoemlb2s.json added

+++ b/reviews/openreview/i8jkjyojy6/notes/5vfoemlb2s.json
@@ -0,0 +1,16 @@
+{
+ "schema": "review/1",
+ "id": "5VFoEMLB2S",
+ "replyTo": "lTSYNSvYHY",
+ "kind": "comment",
+ "invitation": "Official_Comment",
+ "signature": "Authors",
+ "content": {
+ "title": "Supplementary results for L5, Q3, and Q6",
+ "comment": "**Supplement to L5: Threshold selection and transfer**\n\n**Selection.** Each $\\tau_d$ is selected on a held-out route-induction split $\\mathcal{D}_2$ drawn from *training* subjects only and fixed before any test-time evaluation, so it never observes the test distribution. All selected values fall in a narrow band, $\\tau_d \\in [0.9809, 0.9935]$, inside the operative range $[0.98, 1.0]$ swept in Q6.\n\n| Dataset | Selected $\\tau_d$ | Micro-F1 over $\\tau \\in [0.98,1.0]$ | Worst-case loss |\n|---|---|---|---|\n| MIT-BIH | 0.9905 | 0.9498–0.9524 | $\\leq$ 0.3 pt |\n| MIT-BIH Supraventricular | 0.9809 | 0.9509–0.9530 | $\\leq$ 0.2 pt |\n| INCART | 0.9935 | 0.9746–0.9807 | $\\leq$ 0.6 pt |\n| VitalDB | 0.9852 | 0.9116–0.9255 | $\\leq$ 1.4 pt |\n\n**Robustness.** Since every $\\tau_d$ lies inside the swept interval, any threshold in $[0.98, 1.0]$ — including one selected on a different dataset — perturbs Micro-F1 by at most that dataset's own sweep span (last column). Cross-cohort transfer therefore incurs a sub-point cost rather than a failure mode, and the same bound limits the impact of residual mis-calibration.\n\n---\n\n**Supplement to Q3: Calibration protocol and results**\n\n**Protocol.** Confidence is $c = \\max_y p_y$ over the posterior on $\\{\\mathrm{N}, \\mathrm{S}, \\mathrm{V}, \\mathrm{F}\\}$ at the label-generation position, with $\\hat{y} = \\arg\\max_y p_y$; the routing score $C_{\\mathrm{seg}}$ is the mean beat confidence. On the minimal-evidence predictions of each subject-disjoint test split we bin all beats into $M = 10$ equal-width confidence bins and report\n\n$$\\mathrm{ECE} = \\sum_{m=1}^{M} \\frac{|B_m|}{N} \\bigl|\\mathrm{acc}(B_m) - \\mathrm{conf}(B_m)\\bigr|, \\qquad \\mathrm{MCE} = \\max_{m:\\, |B_m| > 0} \\bigl|\\mathrm{acc}(B_m) - \\mathrm{conf}(B_m)\\bigr|,$$\n\nso that sample-weighted ECE reflects a typical beat while MCE reports the worst bin irrespective of size.\n\n| Dataset | ECE | MCE |\n|---|---|---|\n| MIT-BIH | 0.058 | 0.121 |\n| MIT-BIH Supraventricular | 0.029 | 0.117 |\n| INCART | 0.024 | 0.155 |\n| VitalDB | 0.042 | 0.164 |\n\n**Results.** ECE is low throughout ($0.024$–$0.058$): the reliability curve tracks the diagonal at the high-confidence end, where the large majority of beats lie. The larger MCE ($0.117$–$0.164$) is driven by sparsely populated mid-confidence bins with mild over-confidence — benign for routing, which requires only that the score *order* segments by difficulty. This is confirmed by the monotone $C_{\\mathrm{seg}}$–Micro-F1 relationship (Fig. 6) and by rich evidence being acquired predominantly in the lowest-confidence quartile.\n\n---\n\n**Supplement to Q6: Threshold-sensitivity protocol and results**\n\n**Protocol.** For each $\\tau$ we apply the paper's policy unchanged: a segment is answered from the minimal-evidence prediction if $C_{\\mathrm{seg}} \\geq \\tau$ and from the rich-evidence prediction otherwise. The sweep thus traces the full policy family between the pure minimal ($\\tau \\to 0$) and pure rich ($\\tau \\to 1$) models, with endpoints recovering the Minimal- and Rich-specialization rows of Table 1 as verified (MIT-BIH $0.9179 \\to 0.9504$; VitalDB $0.9237 \\to 0.9116$). Since $C_{\\mathrm{seg}}$ concentrates near $1$, essentially no segment reroutes outside $\\tau \\in [0.98, 1.0]$, which we report as the operative range (identical beat set per dataset throughout).\n\n| Dataset | Micro-F1 over $\\tau \\in [0.98,1.0]$ | Span |\n|---|---|---|\n| MIT-BIH | 0.9498–0.9524 | 0.3 pt |\n| MIT-BIH Supraventricular | 0.9509–0.9530 | 0.2 pt |\n| INCART | 0.9746–0.9807 | 0.6 pt |\n| VitalDB | 0.9116–0.9255 | 1.4 pt |\n\n**Results.** Micro-F1 varies by at most $1.4$ points across the operative range, and the curves are smooth — a wide plateau rather than a narrow optimum — so a mis-specified or transferred threshold costs a fraction of a point rather than causing failure. The sweep further shows that the *direction* of benefit is dataset-dependent: rich evidence improves MIT-BIH, MIT-BIH Supraventricular and INCART as $\\tau \\to 1$ but degrades VitalDB, which peaks near the minimal-evidence end. No fixed acquisition policy is therefore optimal across datasets — motivating confidence-gated routing — while the flatness of each curve makes the policy insensitive to the exact threshold."
+ },
+ "createdAt": "2026-07-29T22:03:57.743Z",
+ "modifiedAt": "2026-11-19T16:52:28.469Z",
+ "url": "https://openreview.net/forum?id=i8JkJYOjY6&noteId=5VFoEMLB2S",
+ "license": "CC-BY-4.0"
+}

reviews/openreview/i8jkjyojy6/notes/7paeiw9knz.json added

+++ b/reviews/openreview/i8jkjyojy6/notes/7paeiw9knz.json
@@ -0,0 +1,16 @@
+{
+ "schema": "review/1",
+ "id": "7PAeIw9kNZ",
+ "replyTo": "F1cNFGcVPi",
+ "kind": "comment",
+ "invitation": "Official_Comment",
+ "signature": "Authors",
+ "content": {
+ "title": "Supplementary Results for W1/Q1, W3",
+ "comment": "## Supplement to W1/Q1: Noise robustness\n\n**Setup.** MIT-BIH, evaluated on the same 400-segment subset as the peak-perturbation study (100 all-normal + 100 each containing S/V/F), with the rich-evidence variant (Qwen3.5-4B + frozen ECG-CoCa encoder). All nine conditions share this beat set. The subset is class-enriched, hence the clean reference (Micro-F1 0.8482) sits below the 0.9501 of Table 1 for the same model, while Macro-F1 (0.5609 vs 0.5929) is close.\n\n**Noise.** SNR is 10 log10(signal power / noise power), applied per segment and lead; the noise RMS is about 6%, 13%, 25% and 50% of the signal RMS at 24, 18, 12 and 6 dB. *Gaussian noise* is broadband and corrupts P/T-wave and fine QRS shape; *baseline wander* is low-frequency drift (three random cosines, 0.15–0.5 Hz) that displaces the baseline but preserves morphology. Noise is added to the raw two-lead segment and propagated through the entire pipeline — signal re-encoded, image re-rendered, amplitude/HOS/myMorph re-extracted — so no modality keeps a clean copy; R-peaks and RR are fixed, isolating signal quality from peak-detection error.\n\n| SNR | Gauss. Micro | Gauss. Macro | Wander Micro | Wander Macro |\n| --- | --- | --- | --- | --- |\n| Clean | 0.8482 | 0.5609 | 0.8482 | 0.5609 |\n| 24 dB | 0.8456 | 0.5529 | 0.8480 | 0.5632 |\n| 18 dB | 0.8487 | 0.5638 | 0.8467 | 0.5569 |\n| 12 dB | 0.8409 | 0.5408 | 0.8435 | 0.5478 |\n| 6 dB | 0.8204 | 0.4872 | 0.8431 | 0.5413 |\n\n**Results.** Degradation is graceful, never a collapse. Baseline wander is nearly absorbed (6 dB: Micro -0.5, Macro -2.0 pt) since median-filter baseline removal suppresses drift while morphology survives. Gaussian noise attacks morphology directly: accuracy is unchanged down to 12 dB and drops only at 6 dB (Micro -2.8, Macro -7.4 pt), concentrated on morphology-dependent classes (V 0.848 → 0.755) while the rhythm-anchored N stays stable (0.916 → 0.899). Noise also lowers confidence, so a fixed threshold routes degraded segments to rich evidence more often — the mechanism fails safe.\n\n## Supplement to W3: Cross-dataset evaluation\n\n**Setup.** Four beat-annotated datasets with subject-disjoint splits (MIT-BIH, MIT-BIH-SUP, INCART, VitalDB), AAMI four-class space (VitalDB has no F). For each (source, target) pair the source-trained model is applied to the target's test split; nothing from the target is used for training, tuning, normalization, or threshold selection. For efficiency each pair uses a 10% sample of the target test set (fixed seed) for DeepArrhythmia with threshold-induced routing; the cheaper baselines use the same sample or the full split. DeepArrhythmia keeps its per-dataset checkpoint and routing policy; the 1D ResNet uses the source's normalization statistics; the SVM uses its source-fitted scaler. Diagonals reproduce Table 1, checking the protocol.\n\nDeepArrhythmia (Micro-F1; diagonal = in-domain):\n\n| Train \\ Test | MIT-BIH | MIT-BIH-SUP | INCART | VitalDB |\n| --- | --- | --- | --- | --- |\n| MIT-BIH | **0.9521** | 0.9272 | 0.5540 | 0.5514 |\n| MIT-BIH-SUP | 0.9351 | **0.9527** | 0.8711 | 0.6482 |\n| INCART | 0.9243 | 0.8942 | **0.9791** | 0.6735 |\n| VitalDB | 0.5835 | 0.0935 | 0.7214 | **0.9023** |\n\n1D ResNet:\n\n| Train \\ Test | MIT-BIH | MIT-BIH-SUP | INCART | VitalDB |\n| --- | --- | --- | --- | --- |\n| MIT-BIH | **0.7596** | 0.8911 | 0.8635 | 0.6714 |\n| MIT-BIH-SUP | 0.0743 | **0.8135** | 0.4786 | 0.5827 |\n| INCART | 0.4199 | 0.6787 | **0.8693** | 0.5204 |\n| VitalDB | 0.1012 | 0.0759 | 0.1299 | **0.7859** |\n\nSVM:\n\n| Train \\ Test | MIT-BIH | MIT-BIH-SUP | INCART | VitalDB |\n| --- | --- | --- | --- | --- |\n| MIT-BIH | **0.9366** | 0.8886 | 0.9384 | 0.6735 |\n| MIT-BIH-SUP | 0.8827 | **0.9374** | 0.8624 | 0.6735 |\n| INCART | 0.8825 | 0.8886 | **0.9845** | 0.6735 |\n| VitalDB | 0.8827 | 0.4505 | 0.8624 | **0.7501** |\n\n**Transfer is strong from informative sources.** Trained on MIT-BIH-SUP or INCART — the sources richest in arrhythmia — DeepArrhythmia transfers close to in-domain level (SUP → MIT-BIH 0.9351, → INCART 0.8711; INCART → MIT-BIH 0.9243, → SUP 0.8942). Averaged over these six transfers it reaches 0.824, above the SVM (0.811) and far above the ResNet (0.459), which fails from exactly these sources (SUP → MIT-BIH 0.0743). The SVM's figure is inflated by off-diagonal collapse to majority-class prediction (Macro-F1 0.649 → 0.294), whereas DeepArrhythmia keeps detecting abnormal beats.\n\n**The gap is concentrated on VitalDB.** Transfers into and out of VitalDB account for essentially all degradation (MIT-BIH → VitalDB 0.5514; VitalDB → SUP 0.0935), expected given its different monitor, lead, sampling rate and population; all three methods suffer there. Over the twelve off-diagonal transfers DeepArrhythmia retains mean Micro-F1 0.698 vs 0.457 for the ResNet, so limited transfer reflects domain shift shared by all beat-level methods — and given an informative training distribution the framework generalizes well."
+ },
+ "createdAt": "2026-07-29T22:29:09.634Z",
+ "modifiedAt": "2026-11-19T16:52:28.737Z",
+ "url": "https://openreview.net/forum?id=i8JkJYOjY6&noteId=7PAeIw9kNZ",
+ "license": "CC-BY-4.0"
+}

reviews/openreview/i8jkjyojy6/notes/bld0jsg076.json added

+++ b/reviews/openreview/i8jkjyojy6/notes/bld0jsg076.json
@@ -0,0 +1,16 @@
+{
+ "schema": "review/1",
+ "id": "bld0jSg076",
+ "replyTo": "xaBh41ub8S",
+ "kind": "comment",
+ "invitation": "Official_Comment",
+ "signature": "Authors",
+ "content": {
+ "title": "Supplement to W1: Comparison with Baselines on Cross-Dataset and Rare-Class Generalization",
+ "comment": "# Supplement to W1: Comparison with Baselines on Cross-Dataset and Rare-Class Generalization\n\n**Cross-dataset setting.** Every (source, target) pair is evaluated: the model trained on the source dataset is applied to the target dataset's subject-disjoint test split, giving a full 4×4 matrix. For computational efficiency each pair is evaluated on a 10% sample of the target test set (fixed seed) for DeepArrhythmia with threshold-induced routing; the baselines are far cheaper and are run on the same sample or the full target split. The diagonal is the in-domain setting and reproduces the corresponding Table 1 numbers for all three methods, which validates the protocol.\n\nDeepArrhythmia (Micro-F1; diagonal = in-domain, in bold):\n\n| Train \\ Test | MIT-BIH | MIT-BIH-SUP | INCART | VitalDB |\n| --- | --- | --- | --- | --- |\n| MIT-BIH | **0.9521** | 0.9272 | 0.5540 | 0.5514 |\n| MIT-BIH-SUP | 0.9351 | **0.9527** | 0.8711 | 0.6482 |\n| INCART | 0.9243 | 0.8942 | **0.9791** | 0.6735 |\n| VitalDB | 0.5835 | 0.0935 | 0.7214 | **0.9023** |\n\n1D ResNet:\n\n| Train \\ Test | MIT-BIH | MIT-BIH-SUP | INCART | VitalDB |\n| --- | --- | --- | --- | --- |\n| MIT-BIH | **0.7596** | 0.8911 | 0.8635 | 0.6714 |\n| MIT-BIH-SUP | 0.0743 | **0.8135** | 0.4786 | 0.5827 |\n| INCART | 0.4199 | 0.6787 | **0.8693** | 0.5204 |\n| VitalDB | 0.1012 | 0.0759 | 0.1299 | **0.7859** |\n\nSVM:\n\n| Train \\ Test | MIT-BIH | MIT-BIH-SUP | INCART | VitalDB |\n| --- | --- | --- | --- | --- |\n| MIT-BIH | **0.9366** | 0.8886 | 0.9384 | 0.6735 |\n| MIT-BIH-SUP | 0.8827 | **0.9374** | 0.8624 | 0.6735 |\n| INCART | 0.8825 | 0.8886 | **0.9845** | 0.6735 |\n| VitalDB | 0.8827 | 0.4505 | 0.8624 | **0.7501** |\n\n**Transfer is strong when the source dataset is informative.** What governs transfer is the quality of the training distribution rather than the framework. Trained on MIT-BIH Supraventricular or INCART — the two sources with the richest arrhythmia content — DeepArrhythmia transfers close to in-domain level (MIT-BIH-SUP → MIT-BIH 0.9351, → INCART 0.8711; INCART → MIT-BIH 0.9243, → MIT-BIH-SUP 0.8942). Averaged over the six transfers from these two sources it attains mean Micro-F1 0.824, above the SVM (0.811) and far above the 1D ResNet (0.459), which fails badly from exactly these sources (MIT-BIH-SUP → MIT-BIH 0.0743). The SVM's figure is moreover inflated by its off-diagonal collapse to majority-class prediction (Macro-F1 0.649 → 0.294), whereas DeepArrhythmia keeps detecting abnormal beats. The residual gap is concentrated on transfers into and out of VitalDB, whose monitor, lead configuration, sampling rate and population differ most, and which degrades all three methods; over all twelve off-diagonal transfers DeepArrhythmia retains mean Micro-F1 0.698 versus 0.457 for the 1D ResNet.\n\n**Rare-class setting.** All three methods are evaluated on the identical MIT-BIH DS2 subject-disjoint test split (about 50k beats) with the same AAMI four-class label space. The split is extremely imbalanced — about 44k N, 2k S, 3.2k V and only 388 F beats — which is what makes the minority classes hard for every method.\n\n| Class | DeepArrhythmia | SVM | 1D ResNet |\n| --- | --- | --- | --- |\n| N | 0.974 | 0.967 | 0.858 |\n| S | 0.392 | 0.206 | 0.073 |\n| V | 0.940 | 0.894 | 0.475 |\n| F | 0.066 | 0.025 | 0.040 |\n| **Macro-F1** | **0.593** | **0.523** | **0.362** |\n\n**Rare-class results.** Weak performance on the rare classes is a property of the task, not of our framework: all three methods struggle on S and especially F. Under identical conditions, however, DeepArrhythmia attains the highest F1 on every class, including the two rare ones (S 0.392 vs 0.206 / 0.073; F 0.066 vs 0.025 / 0.040), and the best Macro-F1 (0.593 vs 0.523 / 0.362). Both limitations raised by the reviewer are therefore shared by the baselines, and our framework degrades the least; because its decisions are grounded in explicit peak, rhythm and morphology evidence, its errors also remain inspectable, unlike those of the SVM and ResNet."
+ },
+ "createdAt": "2026-07-29T22:35:28.136Z",
+ "modifiedAt": "2026-11-19T16:52:28.462Z",
+ "url": "https://openreview.net/forum?id=i8JkJYOjY6&noteId=bld0jSg076",
+ "license": "CC-BY-4.0"
+}

reviews/openreview/i8jkjyojy6/notes/jlzrpr8eqw.json added

+++ b/reviews/openreview/i8jkjyojy6/notes/jlzrpr8eqw.json
@@ -0,0 +1,16 @@
+{
+ "schema": "review/1",
+ "id": "jlzRPr8eqW",
+ "replyTo": "VffhuOyVVo",
+ "kind": "comment",
+ "invitation": "Official_Comment",
+ "signature": "Reviewer tYLh",
+ "content": {
+ "title": "Response to authors",
+ "comment": "Thank you for the response to my questions. The proposed method appears to have good noise resilience. I continue to support Accept of this paper."
+ },
+ "createdAt": "2026-08-04T02:39:32.673Z",
+ "modifiedAt": "2026-11-19T16:52:28.324Z",
+ "url": "https://openreview.net/forum?id=i8JkJYOjY6&noteId=jlzRPr8eqW",
+ "license": "CC-BY-4.0"
+}

reviews/openreview/i8jkjyojy6/notes/nhkkkmlyja.json added

+++ b/reviews/openreview/i8jkjyojy6/notes/nhkkkmlyja.json
@@ -0,0 +1,16 @@
+{
+ "schema": "review/1",
+ "id": "NHKKKMlyjA",
+ "replyTo": "Uk8g3kUWhd",
+ "kind": "comment",
+ "invitation": "Official_Comment",
+ "signature": "Authors",
+ "content": {
+ "title": "# Response to the Area Chair (Part 2 of 2)",
+ "comment": "**3. Cross-dataset transfer.**\n\nThe submission already reports a full 4×4 matrix (Appendix I), in which each source-trained model is evaluated on every target test split with nothing from the target used for training, tuning, normalization, or threshold selection (each pair on a 10% sample of the target test set, fixed seed, for computational efficiency). We now apply the identical protocol to the baselines, yielding two findings.\n\nFirst, transfer is strong when the source distribution is informative. Trained on MIT-BIH-SUP or INCART — the sources richest in arrhythmia content — DeepArrhythmia transfers close to in-domain level (MIT-BIH-SUP → MIT-BIH 0.9351, → INCART 0.8711; INCART → MIT-BIH 0.9243, → MIT-BIH-SUP 0.8942), averaging 0.824 over these six transfers against 0.811 for the SVM and 0.459 for the 1D ResNet, which fails precisely from these sources (MIT-BIH-SUP → MIT-BIH 0.0743). Second, over all twelve off-diagonal transfers DeepArrhythmia degrades least (mean Micro-F1 0.698 vs 0.457 for the ResNet), and the failure modes differ in kind: the SVM *systematically* collapses to majority-class prediction off-diagonal, forfeiting abnormal-beat detection entirely (Macro-F1 0.649 → 0.294), whereas DeepArrhythmia continues to detect abnormal beats on most transfers, its worst case confined to the VitalDB-trained direction. Limited transfer is thus attributable to domain shift shared by all beat-level methods — differing sampling rates, leads, devices, and class priors, with VitalDB the principal outlier in both directions — rather than to four independently fitted models.\n\nSeparately, addressing the concern regarding statistical significance, we add 95% bootstrap confidence intervals on in-domain MIT-BIH Macro-F1: DeepArrhythmia 0.593 [0.583, 0.603], SVM 0.523 [0.516, 0.530], and ResNet 0.362 [0.356, 0.367], all non-overlapping.\n\nDeepArrhythmia (Micro-F1; diagonal = in-domain, in bold):\n\n| Train \\ Test | MIT-BIH | MIT-BIH-SUP | INCART | VitalDB |\n| --- | --- | --- | --- | --- |\n| MIT-BIH | **0.9521** | 0.9272 | 0.5540 | 0.5514 |\n| MIT-BIH-SUP | 0.9351 | **0.9527** | 0.8711 | 0.6482 |\n| INCART | 0.9243 | 0.8942 | **0.9791** | 0.6735 |\n| VitalDB | 0.5835 | 0.0935 | 0.7214 | **0.9023** |\n\n1D ResNet:\n\n| Train \\ Test | MIT-BIH | MIT-BIH-SUP | INCART | VitalDB |\n| --- | --- | --- | --- | --- |\n| MIT-BIH | **0.7596** | 0.8911 | 0.8635 | 0.6714 |\n| MIT-BIH-SUP | 0.0743 | **0.8135** | 0.4786 | 0.5827 |\n| INCART | 0.4199 | 0.6787 | **0.8693** | 0.5204 |\n| VitalDB | 0.1012 | 0.0759 | 0.1299 | **0.7859** |\n\nSVM:\n\n| Train \\ Test | MIT-BIH | MIT-BIH-SUP | INCART | VitalDB |\n| --- | --- | --- | --- | --- |\n| MIT-BIH | **0.9366** | 0.8886 | 0.9384 | 0.6735 |\n| MIT-BIH-SUP | 0.8827 | **0.9374** | 0.8624 | 0.6735 |\n| INCART | 0.8825 | 0.8886 | **0.9845** | 0.6735 |\n| VitalDB | 0.8827 | 0.4505 | 0.8624 | **0.7501** |\n\n**4. Role of the Morphology Analyzer.**\n\nIts contribution to accuracy is complementary rather than load-bearing: removed (numerical features only), DeepArrhythmia still exceeds every external baseline (Macro-F1 0.5694 vs 0.523 for the SVM), and the analyzer adds a further 2.4 points. Its principal value is interpretability: whereas the baselines emit only a class label, the analyzer accompanies each decision with clinician-readable, auditable morphological evidence — QRS width, P-wave behaviour, RR regularity — that cardiologists rate as clinically valid (Table 4). This evidence is causal rather than decorative: disrupting the exact cue an explanation references reduces accuracy from 92.5% to 1.9% (Table 13), which likewise addresses the label-leakage concern, alongside split-safe distillation and label-free inference. Such interpretability is essential for decision support and renders even error cases inspectable, which no baseline affords.\n\n**On novelty.**\n\nWe have revised the framing accordingly and do not claim novelty for peak detection, feature extraction, morphology analysis, multimodal fusion, or confidence thresholding in isolation. Our contribution is framework-level and comprises three coupled elements: (1) formulating beat-level classification as peak-aligned prediction within a multi-beat context segment; (2) separating physiological evidence acquisition from multimodal evidence integration through explicit beat-anchor, numerical-feature, and morphology-text interfaces; and (3) training minimal- and rich-evidence agents followed by threshold-induced routed supervised fine-tuning, so that the model learns *when* additional physiological evidence is worth acquiring rather than applying every tool uniformly.\n\nAll new results have been incorporated into the revised appendices, and we would be glad to provide any further analysis the AC or reviewers consider useful."
+ },
+ "createdAt": "2026-07-30T01:11:09.253Z",
+ "modifiedAt": "2026-11-19T16:52:28.373Z",
+ "url": "https://openreview.net/forum?id=i8JkJYOjY6&noteId=NHKKKMlyjA",
+ "license": "CC-BY-4.0"
+}

reviews/openreview/i8jkjyojy6/notes/o53a1uqhbu.json modified

+++ b/reviews/openreview/i8jkjyojy6/notes/o53a1uqhbu.json
@@ -5,11 +5,12 @@
"kind": "meta_review",
"invitation": "Meta_Review",
"signature": "Area Chair UboE",
"content": {
- "metareview": "Four knowledgeable reviewers evaluated this paper, yielding two accept and two borderline accept recommendations. The reviewers are generally positive, viewing the noted weaknesses as addressable limitations rather than barriers to publication. They agree that the paper tackles a clinically meaningful problem with a sound framework, strong motivation, and a thorough empirical evaluation.\n\nKey strengths include a clinically motivated setup treating beat-level arrhythmia classification as a segment-contextualized prediction problem, and an interpretable architecture that decouples physiological measurement from evidence integration. Reviewers also praised the confidence-guided evidence acquisition mechanism for its computational efficiency, alongside comprehensive experiments featuring solid ablations and clear comparisons against non-agentic baselines.\n\nWhile the reviewers noted that the technical novelty stems from integrating existing components rather than introducing a radically new algorithm, they found this to be an acceptable contribution. However, the authors must address specific technical concerns regarding robustness to peak-detection errors and noise, the calibration and threshold stability of the routing mechanism, cross-dataset transfer performance, and the exact role of the Morphology Analyzer.\n\nThe author response should focus heavily on these robustness and generalization questions, particularly routing calibration and cross-dataset behavior. Satisfactory clarifications on these points will keep this paper on its positive trajectory toward acceptance.\n\nNote to Authors: The paper 'A real-time QRS detection algorithm' by Jiapu Pan and Willis J. Tompkins is a seminal paper published in IEEE Transactions on Biomedical Engineering in 1985 (Volume BME-32, Issue 3, pages 230–236). The citation incorrectly lists the publication year as 2007 and the volume as 54. Please fix the discrepancy."
+ "metareview": "Four knowledgeable reviewers evaluated this paper, yielding two accept and two borderline accept recommendations. The reviewers are generally positive, viewing the noted weaknesses as addressable limitations rather than barriers to publication. They agree that the paper tackles a clinically meaningful problem with a sound framework, strong motivation, and a thorough empirical evaluation.\n\nKey strengths include a clinically motivated setup treating beat-level arrhythmia classification as a segment-contextualized prediction problem, and an interpretable architecture that decouples physiological measurement from evidence integration. Reviewers also praised the confidence-guided evidence acquisition mechanism for its computational efficiency, alongside comprehensive experiments featuring solid ablations and clear comparisons against non-agentic baselines.\n\nWhile the reviewers noted that the technical novelty stems from integrating existing components rather than introducing a radically new algorithm, they found this to be an acceptable contribution. However, the authors must address specific technical concerns regarding robustness to peak-detection errors and noise, the calibration and threshold stability of the routing mechanism, cross-dataset transfer performance, and the exact role of the Morphology Analyzer.\n\nThe author response should focus heavily on these robustness and generalization questions, particularly routing calibration and cross-dataset behavior. Satisfactory clarifications on these points will keep this paper on its positive trajectory toward acceptance.\n\nNote to Authors: The paper 'A real-time QRS detection algorithm' by Jiapu Pan and Willis J. Tompkins is a seminal paper published in IEEE Transactions on Biomedical Engineering in 1985 (Volume BME-32, Issue 3, pages 230–236). The citation incorrectly lists the publication year as 2007 and the volume as 54. Please fix the discrepancy.",
+ "final_justification": "The authors addressed the key concerns raised during review through additional experiments and clarifications. The rebuttal improved confidence in the robustness of the Morphology Analyzer, the fairness of the VLM comparisons, and the stability of the confidence-guided routing mechanism. My original reservation remains: the primary contribution is an effective integration of existing components rather than a fundamentally new algorithmic advance. However, the work is technically sound overall, practically relevant, and supported by a sufficient empirical evaluation."
},
"createdAt": "2026-07-20T21:17:24.403Z",
- "modifiedAt": "2026-11-18T15:21:37.036Z",
+ "modifiedAt": "2026-11-19T05:00:59.053Z",
"url": "https://openreview.net/forum?id=i8JkJYOjY6&noteId=o53a1uqhbU",
"license": "CC-BY-4.0"
}

reviews/openreview/i8jkjyojy6/notes/uk8g3kuwhd.json added

+++ b/reviews/openreview/i8jkjyojy6/notes/uk8g3kuwhd.json
@@ -0,0 +1,16 @@
+{
+ "schema": "review/1",
+ "id": "Uk8g3kUWhd",
+ "replyTo": "o53a1uqhbU",
+ "kind": "comment",
+ "invitation": "Official_Comment",
+ "signature": "Authors",
+ "content": {
+ "title": "Response to the Area Chair: Robustness, Calibration, Cross-Dataset Transfer, and the Morphology Analyzer",
+ "comment": "# Response to the Area Chair\n\nWe thank the AC for the clear synthesis and guidance. We have conducted new experiments addressing each of the four highlighted concerns; details appear in our per-reviewer responses and the revised appendices.\n\n**Setup and definitions.**\n\n*Datasets.* All experiments use the four beat-annotated datasets of the paper — MIT-BIH Arrhythmia, MIT-BIH Supraventricular (MIT-BIH-SUP), INCART, and VitalDB — under the AAMI four-class space (N, S, V, F; VitalDB contains no F beats) with subject-disjoint splits. The perturbation and noise studies use a fixed 400-segment MIT-BIH subset (100 all-normal plus 100 each containing S, V, and F beats) so that abnormal beats are adequately represented. \n\n*Models.* DeepArrhythmia comprises a Qwen3.5-4B backbone with a frozen ECG-CoCa signal encoder; the perturbation and noise studies use the rich-evidence variant, the cross-dataset matrix the threshold-induced routed variant. Baselines are the 1D ResNet and the SVM, each trained per dataset on identical splits.\n\n*Metrics.* Micro-F1 is computed over the four classes at beat level; Macro-F1 is the unweighted mean of per-class F1 and therefore reflects the rare classes. ΔF1 = F1_clean − F1_stress on the non-interfered (collateral) beats of a perturbed segment, so positive values denote degradation. ECE is the sample-weighted mean gap between confidence and accuracy over ten equal-width confidence bins.\n\n*Noise and SNR.* SNR is 10 log10(signal power / noise power), applied per segment and per lead; equivalently, the noise RMS is approximately 6%, 13%, 25%, and 50% of the signal RMS at 24, 18, 12, and 6 dB. *Gaussian noise* is i.i.d. and zero-mean, modelling measurement noise and corrupting high-frequency detail (P/T waves, fine QRS shape); *baseline wander* is a low-frequency drift (three cosines at 0.15–0.5 Hz with random phases) modelling respiration and electrode motion, which displaces the isoelectric line while preserving morphology. Each disturbance is injected into the raw two-lead signal and propagated through the entire input pipeline — signal re-encoded, waveform image re-rendered, and amplitude/HOS/myMorph descriptors re-extracted — so that no modality retains a clean copy. R-peak anchors and RR intervals are held fixed, isolating signal quality from peak-detection error.\n\n**1. Robustness to peak-detection errors and noise.**\n\nAll three detector failure modes are now covered: missed and mislocalized peaks were reported previously, and we add *spurious (extra)* peaks. Collateral degradation is graceful in every case and concentrates on the timing-dependent S class (ΔF1: mask 0.019, mislocalize 0.136, extra 0.302), while the normal class remains stable (ΔF1 ≤ 0.028). The noise study shows the same qualitative behaviour: baseline wander is almost entirely absorbed (Micro-F1 −0.5 points at 6 dB), and Gaussian noise is tolerated to 12 dB, declining by 2.8 Micro-F1 points only at 6 dB, with the loss confined to morphology-dependent classes (V-class F1 0.848 → 0.755) while the rhythm-anchored N class holds (0.916 → 0.899).\n\n(a) Collateral ΔF1 under R-peak perturbations:\n\n| Class | Mask | Mislocalize | Extra |\n| --- | --- | --- | --- |\n| N | 0.004 | 0.014 | 0.028 |\n| S | 0.019 | 0.136 | 0.302 |\n| V | 0.043 | 0.025 | 0.072 |\n| F | 0.056 | 0.033 | 0.050 |\n\n(b) Signal-level noise:\n\n| SNR | Gaussian Micro | Gaussian Macro | Wander Micro | Wander Macro |\n| --- | --- | --- | --- | --- |\n| Clean | 0.8482 | 0.5609 | 0.8482 | 0.5609 |\n| 24 dB | 0.8456 | 0.5529 | 0.8480 | 0.5632 |\n| 18 dB | 0.8487 | 0.5638 | 0.8467 | 0.5569 |\n| 12 dB | 0.8409 | 0.5408 | 0.8435 | 0.5478 |\n| 6 dB | 0.8204 | 0.4872 | 0.8431 | 0.5413 |\n\n**2. Routing calibration and threshold stability.**\n\nThe routing score — the mean per-beat maximum class posterior — is well calibrated in-domain on all four datasets, with binned accuracy tracking confidence closely. Performance forms a *wide plateau* in the threshold rather than a narrow optimum: since confidence concentrates near 1, all routing occurs within τ ∈ [0.98, 1.0], and sweeping that entire interval changes Micro-F1 by at most 1.4 points on any dataset, the endpoints recovering the Minimal- and Rich-specialization rows of Table 1 with no discontinuity between them. \n\n| Dataset | ECE | Micro-F1 over τ ∈ [0.98, 1.0] | Span |\n| --- | --- | --- | --- |\n| MIT-BIH | 0.058 | 0.9498–0.9524 | 0.3 pt |\n| MIT-BIH-SUP | 0.029 | 0.9509–0.9530 | 0.2 pt |\n| INCART | 0.024 | 0.9746–0.9807 | 0.6 pt |\n| VitalDB | 0.042 | 0.9116–0.9255 | 1.4 pt |"
+ },
+ "createdAt": "2026-07-30T01:05:38.250Z",
+ "modifiedAt": "2026-11-19T16:52:28.625Z",
+ "url": "https://openreview.net/forum?id=i8JkJYOjY6&noteId=Uk8g3kUWhd",
+ "license": "CC-BY-4.0"
+}

reviews/openreview/i8jkjyojy6/notes/wzsmeu5drd.json added

+++ b/reviews/openreview/i8jkjyojy6/notes/wzsmeu5drd.json
@@ -0,0 +1,16 @@
+{
+ "schema": "review/1",
+ "id": "WzSMeU5dRd",
+ "replyTo": "q5xkQtcxZh",
+ "kind": "comment",
+ "invitation": "Official_Comment",
+ "signature": "Authors",
+ "content": {
+ "title": "Supplement to L: Noise Study — Setup and Results",
+ "comment": "# Supplement to L: Noise study — setup and Results\n\n**Dataset.** The study is run on MIT-BIH, using the same 400-segment evaluation subset as the peak-perturbation study of Appendix O (100 all-normal segments and 100 each containing S, V and F beats; 5,374 beats in total), with the rich-evidence model. All nine conditions (clean plus two noise types at four SNRs) are evaluated on this identical beat set, so differences across conditions are attributable to noise alone.\n\n**Signal-to-noise ratio.** SNR is defined from mean-square powers as 10 log10(signal power / noise power), and the noise is scaled per segment and per lead accordingly. Equivalently, the noise RMS is roughly 6%, 13%, 25% and 50% of the signal RMS at 24, 18, 12 and 6 dB, so 6 dB is a severe corruption.\n\n**Noise models.** *Additive Gaussian noise* is i.i.d. zero-mean Gaussian per sample, modelling thermal/electronic and general measurement noise; being broadband it corrupts high-frequency detail such as P/T waves and fine QRS shape. *Baseline wander* is a low-frequency drift, generated as a sum of three cosines with frequencies drawn uniformly from 0.15–0.5 Hz and random phases, modelling respiration and electrode motion; it displaces the isoelectric line but largely preserves beat morphology. Each disturbance is added to the raw two-lead segment and propagated through the entire input pipeline — the corrupted signal is re-encoded, the waveform image re-rendered from it, and the amplitude/HOS/myMorph descriptors re-extracted from it — so no modality keeps a clean copy. R-peak anchors and RR intervals are held fixed to isolate signal quality from peak-detection error.\n\n| SNR | Gaussian Micro-F1 | Gaussian Macro-F1 | Wander Micro-F1 | Wander Macro-F1 |\n| --- | --- | --- | --- | --- |\n| Clean | 0.8482 | 0.5609 | 0.8482 | 0.5609 |\n| 24 dB | 0.8456 | 0.5529 | 0.8480 | 0.5632 |\n| 18 dB | 0.8487 | 0.5638 | 0.8467 | 0.5569 |\n| 12 dB | 0.8409 | 0.5408 | 0.8435 | 0.5478 |\n| 6 dB | 0.8204 | 0.4872 | 0.8431 | 0.5413 |\n\n**Results.** Degradation is graceful and never a collapse. Baseline wander is almost entirely absorbed (at 6 dB Micro-F1 -0.5 and Macro-F1 -2.0 points). Gaussian noise attacks morphology directly and is harsher, yet accuracy is essentially unchanged down to 12 dB and declines noticeably only at 6 dB (Micro-F1 -2.8, Macro-F1 -7.4 points), with the loss concentrated on morphology-dependent classes (V-class F1 0.848 → 0.755) while the rhythm-anchored N class remains stable (0.916 → 0.899)."
+ },
+ "createdAt": "2026-07-29T22:33:24.410Z",
+ "modifiedAt": "2026-11-19T16:52:28.535Z",
+ "url": "https://openreview.net/forum?id=i8JkJYOjY6&noteId=WzSMeU5dRd",
+ "license": "CC-BY-4.0"
+}

reviews/openreview/i8jkjyojy6/notes/zau88xiqtg.json added

+++ b/reviews/openreview/i8jkjyojy6/notes/zau88xiqtg.json
@@ -0,0 +1,16 @@
+{
+ "schema": "review/1",
+ "id": "ZaU88xiqTG",
+ "replyTo": "7PAeIw9kNZ",
+ "kind": "comment",
+ "invitation": "Official_Comment",
+ "signature": "Reviewer 9zQP",
+ "content": {
+ "title": "response to authors",
+ "comment": "Thank you for the detailed response. I will raise my score as most of my concerns have been adequately addressed."
+ },
+ "createdAt": "2026-08-02T16:56:11.887Z",
+ "modifiedAt": "2026-11-19T16:52:28.332Z",
+ "url": "https://openreview.net/forum?id=i8JkJYOjY6&noteId=ZaU88xiqTG",
+ "license": "CC-BY-4.0"
+}