UltraVoxelGS: Voxel-First Feed-Forward Gaussian Splatting for 3D Ultrasound Reconstruction
Yue Wang, Haoxin Chen, Xiaoqing Guo
NeurIPS 2026
Abstract. We present UltraVoxelGS, a 3D ultrasound reconstruction method that achieves both fast reconstruction and physically faithful rendering for the first time, bringing live spatial feedback during 2D ultrasound scanning within reach. This is made possible by a feed-forward design that predicts a complete Gaussian scene representation from posed ultrasound images directly, entirely bypassing the per-scene optimization bottleneck that has confined prior methods to offline use. Realising such feed-forward prediction in this setting is non-trivial: unlike natural images, ultrasound slices share at most one-dimensional intersections and exhibit view-dependent appearance due to acoustic propagation and attenuation, violating the dense overlap and photometric consistency assumed by existing feed-forward Gaussian methods. Hence, to achieve fast reconstruction, we introduce a feed-forward voxel-to-Gaussian pass that decouples representation size from input count by lifting all posed slices into a fixed-size voxel support and predicting Gaussian parameters directly. For faithful rendering, we propose ultrasound-aware appearance rendering strategy factorize reflection from cumulative attenuation during training and exploit appearance continuity among spatially adjacent slices for efficient Render-time Appearance Adaptation. On both real-world and simulated datasets, UltraVoxelGS achieves best PSNR/SSIM compared with per-scene optimization baselines while reducing inference from 10-20 minutes to approximately 10 seconds—over 35× faster—and scales to sequences exceeding 1,000 frames. Code is available at https://github.com/YuebyYue/UltraVoxelGS-camera-ready.
Are you an author? Claim this paper to see who's reading it, with views and other statistics, and to write about it here.
Claim this paperedit · history
NeurIPS 2026
Accept (poster) · 3 reviews · OpenReview
Updated 25 Nov 2026, 03:44 UTC.
- Decision
- Accept (poster)
Read decision
Meta-review
Summary: the overall initial consensus leans positive, with reviewers finding the core idea technically sound and well-supported. Main strengths: problem formulation is well motivated, proposed method is effective on real and simulated dataset with faster reconstruction times. Main weaknesses: questions regarding the RAA module in evaluation should be further clarified, and lack of discussion on practical clinical use case and dataset consideration.
Final justification
I have read the authors' response and agree with the reviewers' comments that it successfully addressed the issues raised in my initial meta-review. My final recommendation is accept.
Read meta-review
Meta-review
Summary: the overall initial consensus leans positive, with reviewers finding the core idea technically sound and well-supported. Main strengths: problem formulation is well motivated, proposed method is effective on real and simulated dataset with faster reconstruction times. Main weaknesses: questions regarding the RAA module in evaluation should be further clarified, and lack of discussion on practical clinical use case and dataset consideration.
Final justification
I have read the authors' response and agree with the reviewers' comments that it successfully addressed the issues raised in my initial meta-review. My final recommendation is accept.
1 comment
- Rating
- 5/6
- Confidence
- 5/5
- Quality
- 3/4
- Clarity
- 3/4
- Significance
- 3/4
- Originality
- 2/4
- Contribution type
- general
- Ethical concerns
- NO or VERY MINOR ethics concerns only
Read review
Summary
The paper introduces UltraVoxelGS, a 3D ultrasound reconstruction framework that generates fast volumetric reconstructions and physically realistic renderings from 2D ultrasound scans. The method incorporates ultrasound-specific physical properties, including attenuation, reflection, and scattering, into the rendering process. The authors evaluate the method against several baselines and report improved reconstruction/rendering quality, particularly in SSIM, along with substantially faster reconstruction times. Qualitative results also suggest visually impressive 3D reconstructions.
Strengths and weaknesses
This paper addresses an important technical challenge in ultrasound imaging: reconstructing 3D volumetric information from 2D ultrasound acquisitions. The proposed UltraVoxelGS framework is technically interesting because it incorporates ultrasound-specific imaging physics, including attenuation, reflection, and scattering, into the rendering process rather than treating ultrasound as a generic image reconstruction problem. This is a meaningful strength and improves the relevance of the method for ultrasound data.
The manuscript is generally well written and easy to follow. The literature review includes relevant prior work, and the experimental results are promising. The reported improvements in SSIM over the evaluated baselines are significant, and the qualitative results are visually impressive. Another important strength is the substantial reduction in reconstruction time compared with baseline methods, which is highly relevant if the long-term goal is real-time or near-real-time 3D ultrasound visualization.
However, the clinical motivation needs to be clarified. Not every ultrasound application requires 3D reconstruction from 2D scans. In some domains, such as echocardiography, dedicated 3D probes are already available, while in many other applications 2D ultrasound remains the standard clinical workflow. The paper would be stronger if it clearly identified the intended clinical use case. For example, extended field-of-view ultrasound for scoliosis monitoring, long-bone fracture assessment, thyroid imaging, or other anatomy-specific applications could provide clearer motivation for why 3D reconstruction from 2D ultrasound is clinically necessary.
A related limitation is the lack of discussion of existing industry solutions. Commercial approaches already exist for 3D reconstruction from 2D ultrasound acquisitions, such as PIUR Imaging. The paper should explain how UltraVoxelGS differs from these systems, what limitations it overcomes, and why the proposed method would be clinically or technically advantageous. This is difficult to evaluate without a clearer clinical focus.
The description of live 3D reconstruction during acquisition is also unclear. It is not evident how the system can provide a live 3D view before the acquisition is complete. If the goal is sparse acquisition followed by 3D rendering, this should be stated clearly. If instead the method is intended to reconstruct during scanning and guide the operator toward poorly reconstructed regions, then the paper should describe how this feedback is provided. This is especially important because the introduction suggests novice users may struggle with freehand scanning. Without explicit real-time feedback, it is unclear how a novice user would know where to move the probe to improve missing or low-quality regions. The authors should also discuss whether such an interactive workflow could increase total scan time compared with conventional approaches.
A major weakness is that 3D reconstruction accuracy is not directly evaluated. The paper reports image similarity metrics and qualitative visualizations, but it remains unclear whether the reconstructed 3D anatomy is geometrically accurate. For clinical applications, geometric fidelity may be as important as visual realism. The authors should include or discuss metrics that assess 3D spatial accuracy, such as landmark error, surface distance, volumetric overlap, or comparison against ground-truth 3D ultrasound, CT, MRI, or tracked acquisition data where available.
Another important concern is the nature of Dataset B. Since Dataset B is acquired using a robotic ultrasound system and the authors simulate the data by selecting certain frames, this does not represent a true freehand ultrasound acquisition as illustrated in Figure 1. For a proof-of-concept study this may be acceptable, but the manuscript should avoid describing this as “real-world” freehand acquisition unless actual freehand data are used. The absence of validation on true freehand ultrasound data limits confidence in the practical applicability of the method.
Overall, UltraVoxelGS presents a technically promising and visually impressive framework for fast 3D ultrasound reconstruction. Its incorporation of ultrasound imaging physics and its reconstruction speed are important strengths. However, the paper would benefit from a clearer clinical use case, comparison or discussion relative to existing commercial 3D ultrasound reconstruction systems, direct evaluation of 3D geometric accuracy, and validation on true freehand ultrasound acquisitions.
Questions
What is the primary clinical application targeted by UltraVoxelGS? The paper would benefit from a clearer motivation for why 3D reconstruction from 2D ultrasound is needed in the selected use case.
How does UltraVoxelGS differ from existing commercial 2D-to-3D ultrasound reconstruction solutions, such as PIUR Imaging? Please clarify what technical or clinical limitation the proposed method overcomes.
How is live 3D reconstruction intended to work during acquisition? Is the method reconstructing from sparse views after acquisition, or does it provide real-time feedback to guide the operator toward missing or poorly reconstructed regions?
How would a novice user know where to move the probe to improve the reconstruction? If real-time guidance is part of the intended workflow, the paper should describe the feedback mechanism and whether this affects total scan time.
Can the authors evaluate 3D geometric reconstruction accuracy? Metrics such as landmark error, surface distance, volume overlap, or comparison against a tracked/freehand ground truth would strengthen the validation.
Why was the method not validated on true freehand ultrasound data? Since Dataset B is acquired using a robotic ultrasound system and then subsampled, it may not represent the challenges of actual freehand scanning.
Can the authors revise the terminology around Dataset B? Since the data are simulated from robotic acquisition, describing this as real-world freehand ultrasound reconstruction may be misleading.
Limitations
no i have questions and also see the weakness part
Final justification
All my comments have been clearly addressed. The most important inclusion is the addition of the new true freehand ultrasound evaluation (Dataset C). I strongly suggest that the authors include this new experiment in the final camera-ready version. The work will generate important discussions during the meeting. I have upgraded my score to accept
Formatting concerns
no concerns
2 comments
We thank the reviewer for the constructive suggestions and thoughtful comments on our method.
Q1. What is the primary clinical application?
UltraVoxelGS is a general ultrasound reconstruction framework rather than a method designed for a specific anatomical structure. We consider its primary clinical applications to be extended-field-of-view imaging and other static or near-static ultrasound examinations. Representative examples include scoliosis monitoring, long-bone fracture assessment, thyroid imaging, liver, and late-gestation fetal assessment. In these settings, a 3D anatomical representation can provide important diagnostic value, while the native field of view of a dedicated 3D probe may be insufficient to cover the entire region of interest. UltraVoxelGS is not restricted to a fixed 3D probe acquisition volume; instead, its reconstructed field of view can expand with the trajectory of a conventional 2D ultrasound sweep. Moreover, since it requires only a conventional 2D probe, it avoids the additional cost and specialized hardware of dedicated 3D transducers, improving clinical accessibility.
Our experiments demonstrate this capability in two extended-field-of-view settings: Dataset A, consisting of real-world robot-assisted lumbar-spine ultrasound sequences with tracked probe poses, and Dataset B, consisting of simulated liver ultrasound sequences generated from abdominal CT volumes. We will strengthen the clinical motivation in the revised introduction accordingly.
Q2. How does UltraVoxelGS differ from PIUR and other commercial systems?
The key difference lies in the underlying scene representation. UltraVoxelGS uses a direction-aware Gaussian representation, which allows it to model ultrasound-specific appearance factors such as anisotropic reflection and cumulative attenuation. This representation better accommodates flexible scanning trajectories, including local revisits and changes in scanning direction.
In contrast, many conventional or commercial reconstruction systems form an isotropic voxel volume by lifting registered 2D slices into 3D space and compounding their intensities, producing a tomographic-style representation similar to CT or MRI volumes. Such reconstruction commonly favors smooth, approximately uniform sweeps with limited dwelling or repeated sampling of the same region, because redundant observations are often averaged during voxel compounding, which may smooth local details. This can constrain the acquisition trajectory and, consequently, the applicable scanning scenarios.
This difference is particularly relevant near interfaces with large acoustic-impedance differences, where strong specular reflection or acoustic shadowing may hide structures from a single insonification direction. Additional observations from different probe directions can therefore be important. UltraVoxelGS is designed to preserve and render such direction-dependent observations within a unified representation rather than collapsing them into a single isotropic intensity value.
Q3--Q4. How do "live" reconstruction and visual feedback work?
Real-time scan guidance requires two capabilities: (i) fast 3D reconstruction and rendering that keeps pace with the ongoing scan, and (ii) a mechanism to identify under-reconstructed regions and translate them into actionable probe-pose suggestions. This work addresses capability (i). UltraVoxelGS enables a live reconstruction workflow: as the operator scans, the system simultaneously streams acquired frames into a voxel support and, at fixed intervals, generates local GS representations and fuses them into the global scene. This construction proceeds in the background, overlapping with the scan rather than waiting for it. When inspection is needed, the operator can render corresponding views in real time.
Regarding capability (ii), reconstruction incompleteness is already visually observable in the rendered views (see Fig. 11), but it has not yet been converted into explicit guidance. One natural extension is to associate uncertainty with the Gaussian representation and compute a pose-dependent cumulative uncertainty score for candidate poses, directing the operator toward under-observed regions. Beyond visual inspection, the reconstructed scene also supports downstream analysis such as 3D segmentation (see Q5 below), which can further assist the operator in assessing anatomical coverage. These extensions are enabled by, and fundamentally depend on, the fast reconstruction and rendering prerequisite addressed in this work.
Q5. Is the reconstructed 3D anatomy consistent?
We add a 3D anatomical geometry evaluation on Dataset B, where the CT-derived liver segmentation provides the reference anatomy. For each sequence, we use the CT-derived mask on an observed anchor frame as the mask prompt to initialize the same frozen SAM2 model, and propagate the mask to the novel views rendered by each method. The predicted 2D masks are then lifted using their corresponding poses and fused into a 3D anatomical mask and mesh. Because all methods use identical SAM2 initialization and fusion protocol, differences in 3D Dice/IoU are attributable to the geometric and appearance fidelity of each method's rendered views. As shown in Table 1, UltraVoxelGS achieves the best 2D Dice, 3D Dice, and 3D IoU among all evaluated methods, demonstrating superior preservation of 3D anatomical geometry.
This 3D geometry evaluation also directly relates to scan-guidance quality. In practice, scan guidance requires identifying which anatomical regions remain insufficiently covered. Given a more geometrically faithful reconstruction, the operator can reliably compare the reconstructed liver volume against the expected anatomy and localize regions with missing or degraded coverage. UltraVoxelGS therefore enables more trustworthy scan guidance.
Table 1: Segmentation-based 3D anatomical consistency (volume overlap) on Dataset B.
Q6. Validation on true handheld freehand ultrasound.
We first correct the dataset description: Dataset A was acquired using a robotic ultrasound system and then temporally subsampled, whereas Dataset B is simulated liver ultrasound data.
Regarding temporal subsampling, Table 3 already evaluates the effect of decreasing the sampling stride: performance does not degrade as the input becomes denser (the setting contains over 1,600 frames), confirming that subsampling is not masking a difficulty. It is also computationally necessary, since per-scene optimization on the full-density input can exceed one hour, which is impractical for real-world use.
We acknowledge that robotic and handheld acquisitions differ in trajectory characteristics. To directly evaluate the handheld setting, we test the trained model on tracked-handheld Dataset C---paired with Dataset A but differing in acquisition trajectory---without retraining and under the same protocol as Dataset A (Table 2). No baseline is optimized for this handheld setting, yet UltraVoxelGS remains close to its own robotic performance and still outperforms the per-scene UltraGauss, confirming that it generalizes to the more complex trajectories of real-world handheld scanning.
Table 2: Performance on dataset A and dataset C.
- Rating
- 5/6
- Confidence
- 3/5
- Quality
- 3/4
- Clarity
- 3/4
- Significance
- 2/4
- Originality
- 2/4
- Contribution type
- general
- Ethical concerns
- NO or VERY MINOR ethics concerns only
Read review
Summary
This paper proposes a feed-forward Gaussian-splatting framework for 3D freehand ultrasound reconstruction. It lifts all posed 2D slices into a shared voxel support via local inverse-distance interpolation, then predicts voxel-aligned Gaussian primitives from per-target local sub-volumes. Rendering uses a Decoupled Reflection-and-Attenuation (DRA) model (SH reflection × cumulative attenuation), plus a test-time Render-time Appearance Adaptation (RAA) that fine-tunes appearance parameters against the nearest observed slice. On a real robotic lumbar-spine dataset and a simulated liver dataset, it reports the best PSNR/SSIM/LPIPS among the chosen baselines while reducing reconstruction from $\approx$10-20 min to $\approx$10 s.
Strengths and weaknesses
Strengths:
- The paper observes that ultrasound slices share only line-like intersections and are view-dependent, invalidating the assumptions of optical feed-forward GS; decoupling representation size from input count via a shared voxel support is a sensible response.
- The method demonstrates long-sequence scalability (Table 3), and the runtime breakdown (Table 6) is specific. Weakness:
- From a general 3D vision / feed-forward Gaussian Splatting perspective, the core components (voxel-grid interpolation, Transformer-based primitive prediction, and test-time fine-tuning) represent an incremental combination of existing architectures tailored to a specific domain, rather than a broad algorithmic breakthrough for general scene representation.
- As a general feed-forward submission, the paper should include feed-forward 3DGS baselines (e.g., VolSplat) to isolate what the proposed architecture contributes over existing general feed-forward paradigms.
- The headline claim (feed-forward matching per-scene optimization) is carried by RAA, a test-time per-view optimization. Per Table 2, FF+DRA reaches only 19.76 PSNR, below 3DGS (22.55) and UltraGauss (22.06), and RAA alone adds +4.53 dB to 24.29, at 2.63 s/view. This is in tension with the "feed-forward, real-time" framing; Table 1 should report results without RAA.
- "over 60× faster" (abstract) vs "over 10× speedup" (§1), pick one and define the reference.
Questions
- Please report Table 1 results without RAA. Since FF+DRA reaches only 19.76 PSNR (below 3DGS and UltraGauss), how does the pure feed-forward path compare to the optimization baselines?
- RAA is a 50-step per-view optimization at 2.63 s/view. How is the "real-time" claim justified, and how does it differ from a short per-scene optimization?
- Would a more expressive support (ACORN, hash-grid, or TensoRF) preserve the detail lost by voxel interpolation and reduce reliance on RAA?
- Can you add a feed-forward baseline (e.g., VolSplat) to isolate the ultrasound-specific contribution?
Limitations
N/A. It is in the appendix.
Final justification
My major concerns regarding the performance, feed-forward baselines, and ablation study have been successfully resolved. The provided w/o RAA ablation also clarifies the role of appearance adaptation across different pose densities. Overall, the updated paper is technically solid with convincing experiments, justifying a higher score.
Formatting concerns
N/A
4 comments
We thank the reviewer for the detailed questions regarding RAA, spatial support, feed-forward baselines, and runtime.
Q1.Table 1 report w/o RAA and comparison between feed-forward path and optimization baseline
We add UltraVoxelGS w/o RAA (i.e. the feed-forward path with DRA only) to the main comparison table (Table 1). On Dataset B, the feed-forward prediction alone already matches the optimization-based baselines in PSNR; on Dataset A, however, it is noticeably weaker. We attribute this discrepancy to the different pose densities of the two datasets.
On Dataset A, the dense, closely-spaced poses cause interpolation to oversmooth fine structures and degrade the feed-forward prediction; as the dense acquisition also places anchors close to the target, RAA effectively recovers the lost detail, yielding a large gain of about 4 dB. On Dataset B, the sparse poses induce far less smoothing and the feed-forward prediction is already sharp, matching the other baselines; RAA still compensates for the residual interpolation-induced detail loss, contributing a smaller but still substantial gain of about 2 dB..
Overall, the feed-forward pass and RAA are complementary: whenever interpolation smooths out detail, RAA restores it—decisively when the effect is severe (Dataset A) and as a consistent refinement when it is mild (Dataset B). RAA is thus an integral component, not an optional add-on, keeping UltraVoxelGS robust across acquisition densities.
Table 1: Quantitative comparison on real-world and simulated ultrasound datasets.
Q2. Real-time justification and comparison about per-scene optimization
We have optimized the rendering pipeline by removing Gaussians outside the queried slab before rasterization, exploiting ultrasound's thin-slice geometry. The originally reported 2.63 s/view in Table 6 rasterized all Gaussians in the scene; slab-based culling removes over 99.4% of them as irrelevant to the queried slice, reducing rendering time to 2.53,ms/view for direct rendering and 623,ms/view with RAA, without affecting any quality metric. Direct rendering is real-time (2.53 ms); RAA is interactive (623 ms). Both suit the intended clinical workflow where an operator reviews reconstructed slices on demand.
Under this setting, RAA differs from short per-scene optimization in two aspects (shown in Table 2). First, they differ in what is optimized: RAA adapts transient appearance parameters while keeping the reconstructed geometry frozen, whereas per-scene optimization updates the full scene representation itself. Second, they differ in how the cost scales: RAA is a query-time refinement with a fixed per-query cost that is independent of the number of source views, whereas per-scene optimization scales with the length of the input sequence, making it ill-suited to long ultrasound acquisitions. Overall, RAA is not a shortened per-scene reconstruction but a query-time appearance adaptation, offering a more practical refinement strategy for interactive ultrasound inspection.
Table 2: Per-scene optimization versus RAA.
Q3. Discussion about expressive support and RAA
The current voxel support is chosen for simplicity and interpretability. More adaptive representations (ACORN, hash-grid, or TensoRF) are valuable future extensions because their adaptive multi-resolution structures may preserve sharper local details during support construction.
To examine whether a sharper spatial support can reduce the effect of RAA, we increase the voxel resolution at inference time without changing the model or retraining it. As shown in Table 3, the denser support provides a modest improvement for UltraVoxelGS w/o RAA, but increases the Gaussian count from approximately 38.0M to 90.5M. Meanwhile, the complete UltraVoxelGS (w/ RAA) still achieves substantially better performance.
These results suggest that exploring more adaptive and expressive supports is valuable because they can further strengthen the proposed framework. However, stronger spatial support does not replace RAA for two reasons. First, representations such as ACORN or hash grids mainly improve spatial compression and resolution, but do not explicitly address how to preserve direction-dependent appearance under ultrasound anisotropy. Second, consolidating many frame-specific observations into a structured global support inevitably discards some local appearance details unless the compression is entirely lossless. RAA therefore remains necessary to compensate for these lost query-specific details.
Table 3: Expressive support-density analysis.
Q4. Isolate ultrasound-specific contribution
We agree that it is necessary to isolate the ultrasound-specific contribution of UltraVoxelGS using a generic feed-forward GS baseline. However, directly applying VolSplat to long ultrasound sequences is computationally impractical because its native formulation jointly processes a limited number of input views before lifting their features into a 3D feature space and decoding the scene.
We therefore adapt VolSplat to our long-sequence setting while preserving its original training paradigm. During training, randomly sampled source views are used as input, their features are lifted into a 3D feature space, and the complete scene is decoded. During inference, we divide all source views into groups, process each group independently to obtain lifted 3D features, fuse the features from all groups into a common global feature space, and decode the scene representation once. This allows VolSplat to use the full source sequence.
The results on Dataset A are shown in Table 4. UltraVoxelGS w/o RAA substantially outperforms the adapted VolSplat baseline. This gap reflects the distinct reconstruction paradigms. VolSplat must process limited view subsets independently, losing the global scene context and making it difficult to aggregate information distributed across different subset views. In contrast, UltraVoxelGS first consolidates the complete sequence into a unified 3D support and then predicts Gaussians only from the view-interacted local regions. Owing to ultrasound thin-slab locality, each such local region contains the relevant context from the globally constructed support without requiring repeated whole-scene decoding. The result therefore supports the contribution of the ultrasound-specific voxel-first formulation and local-to-global reconstruction strategy.
Table 4: Feed-forward comparison on Dataset A.
Q5. Speedup Clarification
The earlier "over " and "over " statements used different references and were ambiguous. We now use a single, explicitly defined comparison: UltraVoxelGS constructs the global 3D representation in s, approximately faster than the fastest per-scene optimization baseline (ImplicitVol, min).
Thank you for the detailed response and extensive additional experiments. My major concerns are resolved, and please ensure that all these modification and new experimental results are fully incorporated in the revised version. I will raise my score accordingly.
Thank you very much for your constructive feedback and valuable suggestions. We will incorporate the corresponding modifications and additional results into the revised manuscript. We sincerely appreciate your time and effort.
- Rating
- 5/6
- Confidence
- 3/5
- Quality
- 3/4
- Clarity
- 3/4
- Significance
- 3/4
- Originality
- 3/4
- Contribution type
- general
- Ethical concerns
- NO or VERY MINOR ethics concerns only
Read review
Summary
The paper introduces UltraVoxelGS, a method for reconstructing 3D ultrasound from posed 2D ultrasound sequences. It lifts slices into a shared voxel support, predicts voxel-aligned Gaussian primitives from local slab regions, and renders target views with a decoupled reflection/attenuation model. A render-time appearance adaptation step refines transient appearance parameters using a nearby observed slice. Experiments on real robotic ultrasound and simulated liver ultrasound report improved held-out slice reconstruction and faster reconstruction than per-scene optimization baselines.
Strengths and weaknesses
The voxel-first formulation is well motivated by the limited slice overlap and view-dependent appearance of ultrasound. The method combines shared voxel support construction, local Gaussian prediction, and ultrasound-aware rendering in a technically coherent way. The real-data results and long-sequence scalability analysis are practically relevant.
The main concern is that the largest quality gain comes from Render-time Appearance Adaptation: in the ablation, PSNR rises from 19.76 without RAA to 24.29 with RAA on Dataset A. This makes the "feed-forward" and "real-time" framing somewhat dependent on how per-view adaptation cost is counted. The evaluation also focuses on held-out slice reconstruction rather than direct 3D anatomical accuracy, diagnostic utility, or scan-guidance performance. Since the real dataset uses healthy lumbar spine robotic scans, the generality to broader freehand ultrasound, pose noise, device shift, and pathology remains uncertain.
Questions
-
Please report total runtime including RAA for realistic numbers of rendered target views.
-
How does RAA performance vary with distance or angle between the target pose and the nearest observed anchor slice?
-
Please add controlled pose-noise experiments, since the method depends on accurate tracking.
-
Can the authors evaluate 3D geometry or scan-guidance quality beyond held-out slice PSNR/SSIM/LPIPS?
-
For simulated data, can attenuation/reflection components be quantitatively compared with simulator references?
Limitations
The limitations section is useful and covers tracking accuracy, sparse coverage, shadowing, simplified physics, and lack of clinical validation. It should also state more directly that the main evaluation is held-out slice reconstruction and that RAA adds per-view optimization.
Final justification
The rebuttal resolved my main concerns through a clear runtime-quality breakdown, an evaluated streaming implementation, repeated pose-noise experiments, and additional Dataset C and anatomical-consistency results. The authors also appropriately distinguished real-time direct rendering from the slower, higher-quality RAA mode and agreed to revise the claims. RAA still introduces per-view optimization, and clinical validation remains limited, but these limitations are now clearly characterized and are outweighed by the method’s technical contribution and strengthened empirical evidence. I therefore recommend acceptance.
Formatting concerns
Minor language and spacing issues are present.
8 comments
We thank the reviewer for the focused questions regarding RAA, evaluation, and the physical interpretation of the learned components.
Q1. Runtime including RAA
In the intended clinical workflow, reconstruction runs in the background during scanning: our method streams incoming frames into the voxel support and periodically fuses local GS representation into the global scene. The one-time global construction completes in 10.04s, which is well within typical scanning durations, so the representation is ready by the time the operator needs feedback, with no separate reconstruction wait.
Once the global scene is available, the operator-perceived latency per queried view has two modes: (1) direct free-viewpoint rendering at 2.53ms/view (real-time), sufficient for browsing; and (2) RAA-refined rendering at 623,ms/view (sub-second, interactive-rate) for a selected plane requiring physically faithful appearance. (Note: The 2.53ms/view and 623ms/view reflect removing Gaussians outside the queried slab before rasterization, which exploits ultrasound's thin-slice geometry. This does not affect any quality metric; we will update Table 6 accordingly in the revision.) Since construction is fully overlapped with scanning, the latency the operator perceives at inspection time is purely the rendering cost. For RAA-refined views, this is ms , i.e., 6.23s for 10 planes, or 62.3s for 100 views. (Even without overlap, the one-time construction adds only 10s.)
By comparison, per-scene optimization baselines require around 10 minutes even on our subsampled benchmark (approximately 600 frames, corresponding to only approximately 20 s of acquisition at 30 FPS), and their cost grows with input view count (potentially exceeding one hour on denser real-world acquisitions). For background reconstruction to be feasible, its latency must remain shorter than the scanning time; otherwise the representation is still unavailable when feedback is needed. These baselines cannot meet this constraint, making real-time guidance during scanning impractical.
Q2. RAA sensitivity to target--anchor spatial and angular discrepancy.
To analyze the sensitivity of RAA to pose variation, we rank all candidate anchors for each target view by either translational distance or rotational discrepancy, and apply RAA using anchors at different ranks (Top-1, Top-3, Top-5, and Top-10). As shown in Table 1, RAA maintains strong refinement within a small spatial and angular neighborhood, and although the benefit gradually decreases as the anchor's translational or rotational discrepancy grows, it consistently outperforms UltraVoxelGS w/o RAA even at Top-10.
These results support the design motivation of RAA. The feed-forward prediction provides a stable geometric representation, but voxel-interpolation may lead to locally smoothed reconstruction details. Inspired by ultrasound imaging physics, speckle structures preserve local acoustic continuity across neighboring observations. RAA leverages this continuity from nearby anchors to compensate for missing details in the target view while maintaining target-view's geometry. As the spatial or angular discrepancy increases, the local acoustic continuity gradually weakens, leading to reduced refinement benefits.
Table 1: RAA refinement under translation- and rotation-ranked anchors at different Top- ranks on Dataset B.
| Strategy | Anchor | Translation distance (mm) | Rotation angle () | PSNR$\uparrow$ | SSIM$\uparrow$ | LPIPS$\downarrow$ |
|---|---|---|---|---|---|---|
| UltraVoxelGS | Top-1 | 20.87 | 0.3138 | 0.6190 | ||
| Translation | Top-3 | 19.89 | 0.2889 | 0.6919 | ||
| Top-5 | 19.39 | 0.2813 | 0.7018 | |||
| Top-10 | 18.76 | 0.2719 | 0.7136 | |||
| Rotation | Top-3 | 19.91 | 0.2889 | 0.6922 | ||
| Top-5 | 19.41 | 0.2812 | 0.7020 | |||
| Top-10 | 18.68 | 0.2707 | 0.7156 | |||
| UltraVoxelGS w/o RAA | - | - | - | 18.25 | 0.2690 | 0.7860 |
Q3. Controlled pose-noise experiments
The purpose of this experiment is to evaluate whether UltraVoxelGS remains stable when the input poses contain tracking errors. Since training data can be curated with offline pose refinement, we focus on the clinically relevant scenario of inference-time tracking noise. Pose information is used in three stages of our framework: constructing the global representation (fusion), selecting the observed anchor (anchor), and querying the target view (query).
To isolate the sensitivity of each stage, we decompose the effect across four conditions: (A) clean setting; (B) noisy poses are used only for scene fusion; (C) noisy poses are used for both scene fusion and anchor selection; and (D) noisy poses are used for all three stages. Specifically, we inject zero-mean Gaussian noise independently on each translational and rotational axis at two levels ( mm/ and mm/; the latter corresponds to a realized 3-DoF RMS of mm/). As shown in Table 2, fusion is essentially unaffected (B remains at the clean-pose level, within run-to-run variation), and the degradation is concentrated in anchor selection, while target querying adds little. Notably, even under such large perturbations (realized RMS mm/), PSNR drops by only $\sim$1.3 dB, demonstrating that UltraVoxelGS is robust to realistic tracking errors.
Table 2: Controlled pose-noise decomposition on Dataset B.
| Noise | Condition | PSNR$\uparrow$ | SSIM$\uparrow$ | LPIPS$\downarrow$ |
|---|---|---|---|---|
| Clean | A | 20.86 | 0.3158 | 0.6206 |
| 8 mm/4$^\circ$ | B | 20.93 | 0.3176 | 0.6199 |
| C | 20.20 | 0.3033 | 0.6322 | |
| D | 19.86 | 0.2969 | 0.6372 | |
| 12 mm/6$^\circ$ | B | 20.94 | 0.3181 | 0.6200 |
| C | 19.63 | 0.2925 | 0.6419 | |
| D | 19.51 | 0.2900 | 0.6439 |
Q4. Evaluation of 3D geometry
We add a 3D anatomical geometry evaluation on Dataset B, where the CT-derived liver segmentation provides the reference anatomy. For each sequence, we use the CT-derived mask on an observed anchor frame as the mask prompt to initialize the same frozen SAM2 model, and propagate the mask to the novel views rendered by each method. The predicted 2D masks are then lifted using their corresponding poses and fused into a 3D anatomical mask and mesh. Because all methods use identical SAM2 initialization and fusion protocol, differences in 3D Dice/IoU are attributable to the geometric and appearance fidelity of each method's rendered views. As shown in Table 3, UltraVoxelGS achieves the best 2D Dice, 3D Dice, and 3D IoU among all evaluated methods, demonstrating superior preservation of 3D anatomical geometry.
The reconstructed 3D anatomical surface also offers a potential form of scan guidance: the operator can directly inspect whether the accumulated organ boundary is complete or exhibits gaps corresponding to under-scanned regions. A higher 3D Dice/IoU indicates that the reconstruction more faithfully preserves the true organ geometry, making such coverage gaps easier to identify and localize. UltraVoxelGS achieves the best scores on both metrics, suggesting it provides the most reliable geometric basis for this type of volumetric coverage feedback.
Table 3: Segmentation-based 3D anatomical consistency on Dataset B.
| Method | 2D Dice$\uparrow$ | 3D Dice$\uparrow$ | 3D IoU$\uparrow$ |
|---|---|---|---|
| UltraVoxelGS | 0.8469 | 0.8787 | 0.7866 |
| 3DGS | 0.7441 | 0.8593 | 0.7557 |
| ImplicitVol | 0.6419 | 0.7283 | 0.5758 |
| UltraGauss | 0.5153 | 0.6802 | 0.5188 |
| Kernel interpolation | 0.6937 | 0.7683 | 0.6277 |
| Ultra-NeRF | 0.5215 | 0.5776 | 0.4115 |
Q5. Quantitative evaluation of learned physical components
We quantitatively compare the learned attenuation factor with the simulator attenuation reference on Dataset B: . Compared with a depth-linear decay baseline, whose slope and intercept are least-squares fitted to the reference (, ), our results confirm that the factorization captures non-trivial attenuation structure despite receiving no supervision.
The simulator does not provide an independent reference corresponding to the reflection component; Nonetheless, the qualitative visualization in Fig. 7 shows that the reflection component captures boundary structures consistent with tissue interfaces.
Q6. generality to broader freehand ultrasound,
We directly test our Dataset A–trained model on Dataset C, a paired handheld freehand dataset with a different acquisition trajectory. Without retraining, UltraVoxelGS still remains close to its own robotic performance (see cGgJ.Q6), indicating robustness to trajectory change and pose noise.
Thank you for the additional experiments; the anchor, pose-noise, anatomical-consistency, and attenuation analyses strengthen the evaluation. My main remaining concern is the runtime framing. The submitted appendix reports 0.18 s/view without RAA and 2.63 s/view with RAA, whereas the rebuttal reports 2.53 ms and 623 ms. Please explain the implementation, hardware, or timing-protocol changes and pair each runtime with its corresponding quality. The submitted ablation shows 19.87 PSNR without RAA versus 24.29 with RAA, so the highest reported quality is not obtained by the direct real-time mode. Please also clarify whether the claimed streaming/periodic reconstruction was implemented and evaluated or is only a proposed workflow, since the submitted method describes constructing the support from the input sequence. For pose noise, specify whether perturbations are independent per frame and how many realizations were used. The SAM2 experiment is useful but should be described as segmentation-based anatomical consistency rather than a direct measurement of Gaussian geometry. Please also include the actual Dataset C numbers. These clarifications will affect my final assessment.
From OpenReview, CC BY 4.0. Authors as signed there; shown unedited.
Loading the discussion…
We thank the AC for the constructive summary. Below we address the two flagged concerns and highlight additional experiments conducted during the rebuttal.
First, regarding the role of RAA, our method supports two complementary modes: a direct mode for real-time browsing during scanning, and an RAA mode for sub-second, high-fidelity inspection of selected views. RAA complements rather than replaces the feed-forward reconstruction, compensating for residual details lost in voxel interpolation. As the two modes target different operating points (i.e., real-time browsing versus highest-fidelity inspection), we will make this distinction explicit in the revision. (see 21lz.Q1 and wZmx.Q1-Q3).
Second, regarding clinical applicability and dataset considerations, the primary motivation is extended-field 3D ultrasound imaging, and unlike existing commercial solutions, our direction-aware representation preserves view-dependent acoustic appearance for more flexible scanning. To further validate applicability under real-world deployment conditions, we provide freehand dataset evaluation (Dataset C) and controlled pose-noise experiments, demonstrating robustness under more realistic acquisition (see cGgJ.Q1-Q2, cGgJ.Q6-Q7, and 21lz.Q3).
Additionally, following reviewers' suggestions, we strengthen the paper in several directions. Regarding the intended workflow and real-time capability, reconstruction runs in the background during scanning: incoming frames are streamed into the voxel support and periodically fused into the global scene, making the reconstructed representation available in real time as scanning proceeds, so the operator can obtain feedback without any separate reconstruction step. We also implement and evaluate a streaming reconstruction variant demonstrating this workflow, and resolve a runtime discrepancy noted during review by applying a rendering-stage optimization that accelerates inference while leaving all quality metrics unchanged (see 21lz.Q1 and cGgJ.Q3-Q4). Regarding volumetric reconstruction quality, we additionally supplement a segmentation-based 3D anatomical consistency evaluation, demonstrating the superior anatomical geometry preservation of our method over baselines.(see 21lz.Q4 and cGgJ.Q5). Regarding the ultrasound-specific contribution, we additionally include a generic (physics-agnostic) baseline, demonstrating the benefit of our ultrasound physics modeling (see wZmx.Q4).
We hope these clarifications and experiments address the AC's and reviewers' concerns and strengthen our work.